1. Touch

SOUR on the main mode screen.

2. Touch setup to set parameters essential for the SOUR

calculation. These include dilution ratio, minimum run time,
maximum run time, minimum DO required to start the test, the
minimum DO required to end the test and the total solids of the
sample in g/L.

3. Touch

start to initiate the SOUR test.

4. Touch

stop to terminate the test at any time.


A timer on the SOUR screen will display the time of the test. When
the test concludes, the SOUR results in mg/L/hr will be displayed.
Touch print to save the data into the data log, or touch OK to return
to the SOUR screen.

The calculation employed for the SOUR result is:

SOUR= OUR / solids weight (g/L) = mg/hr/g