The DO 5000 has an internal barometer. When it is no longer reflects
the true barometer pressure, it must be calibrated. The Set
Barometer option allows you to do this.

To Set Barometer

1. Access the Set Barometer screen from the DO/ BOD/ OUR/

SOUR setup screen. The current pressure is displayed on the
screen


2. Choose the pressure units desired by touching one of the two

unit buttons, mm Hg or mbars.


3. Use

the

keypad to enter the local barometric pressure, obtained

from a barometer. Do not use weather bureau pressure, as it is
corrected to sea level.


4. Touch

enter to accept the reading and exit to the Setup screen.

OR

Touch exit to return to the DO/ BOD/ OUR/ SOUR Setup screen,
without making a change.