Thermo Fisher Scientific CyberScan DO 5000 User Manual
Page 59
Instruction Manual
CyberScan DO5000
56
2. Touch
MAN to set the meter for manual printing. Touch enter to
accept the print interval mode and return to the Setup screen.
Printing is now done manually by touching print on the Measure
screen.
OR
Touch exit to return to the Setup screen, without making any
changes.
For Stable Reading Printed
In this mode, data is printed every time the meter recognises the
current measurement as stable.
1. Access the Set Print Interval screen from the DO/ BOD/ OUR/
SOUR Setup screen. The current print interval is displayed on the
screen.
2. Touch
stable to set the meter for stable reading printing.
3. Touch
enter to accept the print interval mode and return to the
Setup screen.
Printing is now done when the meter recognises the present
reading as stable.
OR
Touch exit to return to the Setup screen, without making any
changes.
For Timed Interval Printing
In this mode, data is printed at the timed interval that you select.
1. Access the Set Print Interval screen from the DO/ BOD/ OUR/
SOUR Setup screen. The current print interval is displayed on the
screen.
2. Touch
timed to access the timed interval mode and delete the
current print interval time.
3. Use the keypad to enter the desired time for the print interval.
4. Touch enter to accept the new time interval for printing and
return to the Setup screen.
Printing is now done at the set timed interval.
OR
Touch exit to return to the Setup screen, without making any
changes.