Thermo Fisher Scientific CyberScan DO 5000 User Manual

Page 84

background image

Instruction Manual

CyberScan DO5000

81

8. Key in the volume of the SEED sample using the alphanumeric

touch pad. Touch enter to accept the SEED volume.

STABLE

Not Standardized

SEED volume

1 2 3

4 5 6

7 8 9

- 0 .

BS exit enter

123

9. Touch

enter to accept the Seed dilution ratio.

10. Use the bottle volume touch pad to key in the mL volume of the

BOD bottle.

STABLE

Not Standardized

Bottle volume

1 2 3

4 5 6

7 8 9

- 0 .

BS exit enter

123

11. Touch enter to accept the bottle volume and return to the BOD

measure screen.