Instruction Manual

CyberScan DO5000

41

5.3

Select Calibration Mode




















This mode permits a choice between automatic and manual
standardization. In most case, automatic calibration is preferred.
However, if the standard is not saturated and the oxygen
concentration is known, (by titration, for example), manual calibration
offers the user capability of entering the known oxygen concentration
via a numeric keypad.

1. Access the Select Standardization Mode from the Setup screen.

The current mode is shown on the screen.


2. Touch

AUTO or MAN to choose the desired mode.


3. Touch

enter to accept the choice and return to the setup screen.

Current MODE

Select D.O. Calibration Mode

enter

AUTO

MAN