background image

Installation and operating instructions

25

BATCHFLUX


4.5 Error messages in measuring mode


The following list gives all errors that are likely
to occur during process flow measurement.
Errors are shown in the display when “YES“ is
set in Fct. 1.4 DISPLAY, subfunction
“DISP.MSG.“.

Error message

Discription of error

Error clearance

LINE INT.

Power failure
Note:
no counting during power failure

Delete error message in RESET/QUIT
menu. If necessary, reset totalizer.

PULS.OUTP. P

Pulse output overranged.
Note:
totalizer deviation possible.

Check instrument parameters and
correct, if necessary. After cause has
been eliminated, error message is
deleted automatically.

ADC

Analog/digital converter
overranged.

After cause has been eliminated, error
message is deleted automatically.

FATAL. ERROR

Fatal Error; all outputs set to
“Min. values“.

Please consult factory.

TOTALIZER

Totalizer has been reset.

Delete error message in RESET/QUIT
menu.

4