Error messages in measuring mode

Warnings

Description of error

Eliminate error

Pipe (partially) empty

Fill pipe

Poor grounding, or none at all

Check grounding system

Electrical conductivity too low

Check process liquid

Electrodes contaminated

Clean electrodes

Pipe empty (1)

Electrode wire break

Repair wires

Field coil defective

Short-circuit, break, or excess
temperature

Check and eliminate fault

Analog/digital converter (ADC)
defective

Replace electronic unit

Primary head defective

Replace primary head

K 50 constant incorrect

Correct, see nameplate

Linearity

ADC zero incorrect

Replace electronic unit

Low energy

Energy too low for correct
measurements

Voltage is less than 14 V,
increase energy

Overranging (2)

ADC overranged

Change Fct. 3.4 Limitation

Overflow current

Measured value higher than full-
scale range

Check device parameters and
correct if necessary

Overflow pulse (3)

Pulse rate too high,
max. 1 / (1.5×pulse width)

Check device parameters and
correct if necessary

Overflow counter

Overflow of a counter

Reset counter

Line interrupt (4)

Power failure

Delete error message and, if
necessary, reset counter

Fatal error

Severe error, measurement
interrupted

Replace electronic unit

(1)

Is checked only when “Application“ is activated in Fct. 3.4

(2)

Value for checking overranging is set in Fct. 3.4 "Application" relative to full-scale range

(3)

Missing pulses are recovered at times when pulse rate is low

(4)

Is checked only when a counter is activated in Fct. 1.4 "Display"

4.6

Reset counters and cancel error messages

Display texts

Description and settings

... measuring mode ...

Entry into RESET menu

FCt. 4.0

RESET MENUE

FCt. 4.1

ERROR RESET

Resetting the messages
power failure and counter overflow

rESEt

NO

YES

FCt. 4.2

COUNTER RESET

Reset
all counters

rESEt

NO

YES