4.5
Error messages in measuring mode
Warnings
Description of error
Eliminate error
Pipe (partially) empty
Fill pipe
Poor grounding, or none at all
Check grounding system
Electrical conductivity too low
Check process liquid
Electrodes contaminated
Clean electrodes
Pipe empty (1)
Electrode wire break
Repair wires
Field coil defective
Short-circuit, break, or excess
temperature
Check and eliminate fault
Analog/digital converter (ADC)
defective
Replace electronic unit
Primary head defective
Replace primary head
K 50 constant incorrect
Correct, see nameplate
Linearity
ADC zero incorrect
Replace electronic unit
Low energy
Energy too low for correct
measurements
Voltage is less than 14 V,
increase energy
Overranging (2)
ADC overranged
Change Fct. 3.4 Limitation
Overflow current
Measured value higher than full-
scale range
Check device parameters and
correct if necessary
Overflow pulse (3)
Pulse rate too high,
max. 1 / (1.5×pulse width)
Check device parameters and
correct if necessary
Overflow counter
Overflow of a counter
Reset counter
Line interrupt (4)
Power failure
Delete error message and, if
necessary, reset counter
Fatal error
Severe error, measurement
interrupted
Replace electronic unit
(1)
Is checked only when “Application“ is activated in Fct. 3.4
(2)
Value for checking overranging is set in Fct. 3.4 "Application" relative to full-scale range
(3)
Missing pulses are recovered at times when pulse rate is low
(4)
Is checked only when a counter is activated in Fct. 1.4 "Display"
4.6
Reset counters and cancel error messages
Display texts
Description and settings
... measuring mode ...
↵
Entry into RESET menu
FCt. 4.0
RESET MENUE
→
FCt. 4.1
ERROR RESET
↑
Resetting the messages
power failure and counter overflow
rESEt
NO
↑
YES
↵
FCt. 4.2
COUNTER RESET
→
Reset
all counters
rESEt
NO
↑
YES
↵