beautypg.com

KROHNE ALTOFLUX 2W IFM 4042K EN User Manual

Page 22

background image

Section 4.4

Part B IFC 040 Signal converter

ALTOFLUX 2W

02/2001

22

4.4

Table of settable functions

Display texts

Description and settings

FCt. 1.1

FULL SCALE

Full-scale range

XXX.XXX

m3/h

l/s

Ga./m

"user unit"

Range:

0.3-12 m/s = 1-40 ft/

s

Set full-scale range, i.e. for the maximum occurring flow rate.
This affects all functions where values have to be set as % of
full-scale range:

Fct. 1.3 SMU

Fct. 1.5 Current output

Fct. 1.7 Status output

Fct. 2.1 Test Q

Fkt. 3.4 Application

FCt. 1.2

TIME CONSTANT

Time constant

XX.X

S

Range 0.5 ... 99.9s

Standard 3.0s

Setting the time constant for a two-pole low pass for signal
damping. This time is needed after an abrupt change in the
flow rate in order to attain 67% of the new flow value.
The time constant acts equally on the current output and the
display, and also on the status output when the ‘sign’ or ‘limit
value’ functions are selected.
The value can be transmitted via the HART interface with or
without use of the time constant.
If the time constant is changed, the noise level needs to be
reset if the noise filter is activated (see Fct. 3.4 Application
“Filter setting").

FCt. 1.3

L.F. CUTOFF

Low-flow cutoff (SMU)
For characteristic, see Fig. 1 in Section 2.3

XX XX

Perc.

Range 1 ... 20%

(1

st

value < 2

nd

value)

Standard: 04 ... 05%.

At low flows, flow display and counting are suppressed. This
also applies to the negative flow direction.
Cutoff ‘off’ value (1st figure) and cutoff ‘on’ value (2nd figure)
must be set as a percentage of the full-scale range (see Fct.
1.1 Full-Scale). The SMU acts on the current output, pulse or
status output, all counters and also values via the HART
interface and the display.

FCt. 1.4

DISPLAY

Display
Setting for display form of measured values and messages in
the local display. The following settings are possible:

inFo

diSPlaY Flow

Display of flow

Edit

Percent

no diSPlaY

m3/h

l/s

Ga./m

"user unit"

Standard: Percent

Select unit for display of current flow.
Nothing displayed when “no display" has been set.