Section 1.7

Part A Installation and Start-up

ALTOFLUX 2W

02/2001

10

1.7

Grounding

All flowmeters must be properly grounded to avoid personnel shock hazard.

The ground conductor should not transmit any interference voltages, therefore do not
ground any other electrical devices together with this conductor.
Repeater power supply unit

Protective separation (PELV) must be ensured
(VDE 0100 / VDE 0106 or IEC 364 / IEC 536, or equivalent national regulations).

For measurement reasons, connect an FE functional ground conductor.

Metal pipelines, not internally coated

grounding without grounding rings

Metal pipelines,

with or without internal coating,

and plastic pipelines

grounding with grounding rings

D1, D2, D3

Gaskets, not included with supply, to be provided by customer.

E

Grounding rings (option)

F

Flowmeter flanges

FE

Functional ground, wire

4 mm

2

Cu (10 AWG), not included with supply,

to be provided by customer.

R

Pipeline

RF

Pipe flanges

V1, V2

Interconnecting wires, included with supply

Y

Terminal box or signal converter