Part B IFC 040 Signal converter

Section 4.4

02/2001

ALTOFLUX 2W

25

Display texts

Description and settings

FCt. 1.6

PULSE OUTPUT

Pulse output
Settings for the pulse output

Characteristic, see Fig. 3 in Section 2.3

Only possible when “Pulse output” is set under Fct. 3.5
Hardware.
When “Status output” selected, Fct. 1.6 has “no function".

inFo

Function

Function

Edit

off

PoS.direction
neG.direction

abSolute

Standard: absolute

Setting of the characteristic of the pulse output
“off” = switch at output open

Other selection

inFo

PulSe width

Pulse width
Minimum interpulse period = half pulse width

XXX0.

mS

Range 30 ... 1000 ms

Standard: 50 ms

Pulse width defines the time during which the switch at the
output is closed and high current flows between terminals B1
or B2 and B

. The maximum pulse rate is selected at the

same time, as the interpulse period is at least equal to half

the pulse width:

width

pulse

1.5

1

rate

Pulse

max

×

=

inFo

Pulse/Volume

Pulses / Volume
Number of pulses per unit volume

XXX.XXX

m3

l

Gal.

“user unit"

Range 0 ... 10 Hz

Standard:

1 pulse per m3

Pulses/Volume is used to set the number of pulses that are
output for the given volume. If 10.0 is set at Unit m

3

, 10

pulses are output per cubic metre. If 0.01 is set at Unit l , one
pulse is output per 100 litres.

A large pulse width together with a high pulse rate will
cause overranging. Therefore, pulse rate is limited so
that the minimum interpulse period does not drop below
half the pulse width. In that case, error message due to
overranging of the pulse output, i.e. marker flashes and,
if activated in Fct. 1.4 “Display", is output in the form of
horizontal scrolling.

When the pulse output is overranged, the missing pulses
are output later, at times of lower flow.