KROHNE ALTOFLUX 2W IFM 4042K EN User Manual
Page 24
Section 4.4
Part B IFC 040 Signal converter
ALTOFLUX 2W
02/2001
24
Display texts
Description and settings
FCt. 1.5
CURRENT OUTPUT
→
Current output
For characteristic, see Fig. 2 in Section 2.3
For HART function “No" or
address “0"
Settings for the current output
Not possible when ”Address 1 –15” is set under Fct. 3.6
HART (equivalent to multidrop mode).
In that case, only a constant current needs to be set,
see Fct. 3.6 HART “I Multidrop".
Fct. 1.5 Current output then has “no function".
inFo
Function
↵
Edit
off
PoS.direction
neG.direction
abSolute
↑
neG-0-PoS
Standard: absolute
↵
Setting the characteristic of the current output
“off” = 0 per cent current output
Please note:
Fct. 1.7 Status output, “Automatic range change" setting.
inFo
0 Perc.current
↵
0 per cent current
XX.X
mA
Range 4.0 ... 14.0mA
Standard: 4.0 mA
↵
Setting of current at “zero” flow(I
0%
)
inFo
100 Perc.current
↵
100 per cent current
XX.X
mA
Range 10.0 ... 20.0 mA
I
0%
< I
100%
Standard: 20 mA
↵
Setting of current at 100 % flow (I
100%
),
according to the full-scale range (Q
100%
)
under Fct. 1.1 Full Scale.
inFo
error current
↵
Fault current
XX.X
mA
Range 3.6 ... 22.4 mA
I
Error
< I
0%
or I
100%
< I
Error
Standard: 22 mA
↵
The fault current that is output in the event of a fault.
Please note:
When the current output is overranged, the maximum current
is 21mA and has been preset by factory.