TECHNICAL DATA

52

BM 702 A

www.krohne.com

09/2009 - 4000643301 - MA BM 702 A R01 en

Process connections

Horn antenna or Wave-Guide

DN50...200, PN6...64

ASME B16.5: 2...8", 150/300 lbs/RF

Wave-Stick

DN50...150

ASME B16.5: 2...6", G1½, 1½NPT

Dairy screw connection

DIN 11851: DN50, DN65, DN80

SMS 1145: 51 mm, 63 mm, 76 mm

Tri-clamp

ISO 2852: 2...4"

Electrical connections

Power supply

Non-Ex / Ex i: 14.5…30 VDC

Cable entry

M20x1.5; ½NPT

Cable gland

Standard: M20x1.5

Options: ½NPT, G½ adapter

Cable entry capacity (terminal)

0.5…1.5 mm

2

/ AWG 20...16

U-clamp terminals

(for PA and FE)

Max. 4 mm

2

/ AWG 12

Output

Description of the used

abbreviations

U

ext

= external voltage; R

L

= load + resistance

Current output

Current output

Current output

Current output

Output data

Distance, level and volume

Measuring range

4…20 mA HART

®

or 3.8…20.5 mA acc. to NAMUR NE 43

1

Load

Minimal 0 Ω; max. R

L

= ((U

ext.

- 14.5 VDC) / 22 mA)

Temperature drift

≤ 150 ppm/K

Error signal

Acc. to NAMUR NE 43

High value: 22 mA

Low value: 3.6 mA

HART

HART

HART

HART

®

Description

HART

®

protocol via current output

HART

®

version: V5

Load

≥ 250 Ω at HART

®

test point;

Note maximum load for current output!

Multidrop operation

Yes, current output = 4 mA

Multidrop address adjustable in operation menu 1…15

.book Page 52 Thursday, September 3, 2009 2:53 PM