3.6.3 Bypass chambers

Install a bypass chamber next to the tank if:

• There is highly conductive foam in the tank.
• The liquid is very turbulent or agitated.
• There are too many obstacles in the tank.

Figure 3-9: Installation recommendations for bypass chambers

1 Bypass chamber
2 Clearance between the antenna and the wall of the stilling well ≤ 5 mm / 0.2"
3 Sudden change in well diameter ≤ 1 mm / 0.04"

CAUTION!

Installation requirements

The bypass chamber must be electrically conductive.

The inside diameter of the bypass chamber must not be more than 10 mm / 0.4" over the
diameter of the antenna.

The bypass chamber must be straight.

The bypass chamber must have a surface roughness of ±≤0.1 mm / 0.004".

There must be no sudden changes in internal diameter greater than 1 mm / 0.04".