6.8 Sensor date submenu
In the sensor data submenu you have the opportunity to activate sensor-specific functions. You can offset
various sensors with one another and thus actively reduce cross-sensitivities from unknown ions.
You get to the sensor data submenu via the measuring channel menu.
Settings for disturbing ion compensation
Designation
Settings
Description
Channel number
00
Set the other channel that is to be used for compensation here.
Coefficient
-9.99 - +9.99
Offsetting factor by which the cross-sensitivity in respect of counter-ions is to be
compensated. Please ask KROHNE for the specific values.
Addition
Yes
No
Yes
Activation of compensation
No
Deactivation of compensation
Electrode specification (factory setting)
Designation
Settings
Description
Inner buffer
+- 000 mV
Sensor-specific features. Loaded on retrieval of the sensor. Please do not change the
factory setting.
Norm-Offset
+- 0.00 pX
Sensor-specific features. Loaded on retrieval of the sensor. Please do not change the
factory setting.