6.8 Sensor date submenu

In the sensor data submenu you have the opportunity to activate sensor-specific functions. You can offset

various sensors with one another and thus actively reduce cross-sensitivities from unknown ions.

You get to the sensor data submenu via the measuring channel menu.

Settings for disturbing ion compensation

Designation

Settings

Description

Channel number

00

Set the other channel that is to be used for compensation here.

Coefficient

-9.99 - +9.99

Offsetting factor by which the cross-sensitivity in respect of counter-ions is to be
compensated. Please ask KROHNE for the specific values.

Addition

Yes

No

Yes
Activation of compensation

No
Deactivation of compensation

Electrode specification (factory setting)

Designation

Settings

Description

Inner buffer

+- 000 mV

Sensor-specific features. Loaded on retrieval of the sensor. Please do not change the
factory setting.

Norm-Offset

+- 0.00 pX

Sensor-specific features. Loaded on retrieval of the sensor. Please do not change the
factory setting.