Settings for auto-clean function winter service (necessary when rinsing with water)

Designation

Settings

Description

Interval

000.0 h

Entry at which time intervals the auto-clean function is to be activated.
Recommendation: 000.2 h

Duration

000 sec

Rinse duration in seconds

Recommendation: 002 sec.

Hold time

000 sec

Waiting period after rinsing after which measurement can be engaged again. The
analogue outputs are frozen during the hold time.

Temperature

008 °C

Limit temperature for winter service program.

Automatic

Yes

No

Yes

The auto-clean function is active

No

The auto-clean function is not active.

Switch to other menu levels

Designation

Settings

Description

Menu

MainMnu

CalMnu

SensorMnu

MainMnu

Return to the main menu

CalMnu

Call up the calibration menu

SensorMnu

Call up the specific sensor menu