6.7 Calibration submenu
You set the specific parameters for the calibration routines in this menu.
You get to the calibration submenu via the measuring channel menu.
Entering current value for offset calibration
Designation
Settings
Description
Current pt
00.00 ppm
Entry of the laboratory value for offset calibration
Settings for standard addition (optional)
Designation
Settings
Description
Sample volume
10.00 lt
Pre-set volume in which calibration is carried out following standard addition.
Flow
000.02 ml
Inflow rate of the standard calibration solution
Standard
0.000 M/l
Concentration of the standard in Mol/l
No. steps
03 step
Enter number of steps with which the standard addition is to be implemented.
2-point calibration: Entering the concentrations of the standard solutions
Designation
Settings
Description
1st buffer
000.0 ppm
Entry of the concentration of the first calibration standard (low concentration)
2nd buffer
000.0 ppm
Entry of the concentration of the second calibration standard (higher concentration)