5.7 The most important settings for a quick start

You can quickly adapt the measuring system to your needs if you pay attention to the small number of

settings described below. All further changes and configurations can be carried out at any time during

ongoing operation.

The measuring system is pre-configured factory-side in accordance with the order data.

Electrode type

Check whether the set electrode type matches the electrode installed in the immersion holder in the

respective measuring channel (= Addr.No.). Pay attention to electrode identification and plug no. (=

Addr.No.). The electrode type (measuring parameter) is displayed top right in the called-up measuring

channel.

Change the electrode type where necessary under "Menu – Sensor data – Electrode".

Measuring range = mA-output

Set the mA range that you need for the respective sensor:

"Menu – Power output – mARange / Start point / Finish point"

Example: Ammonium (NH4-N)

0 .. 20 mA ----- 0 .. 50 mg/l NH4-N