Settings for the display

Designation

Settings

Description

Units

lin

log

mV

pH

%

The selection possibilities depend on the respective parameter. The reference
magnitude between molecule or element can also be chosen during ISE measurement.

Modify

permitted

blocked

save data

load data

permitted
Entries and modifications are possible for this channel.

blocked
Entries and modifications are not possible

save data
Saves the current measuring channel settings.

load data
loads and uses the last saved channel settings.

System

Reset

Default

Test

Reset
The converter is restarted with the currently used parameters

Default
The converter is restarted with the factory settings. Existing customised settings are
lost.

Test
The converter switches to a test mode. Use this setting only in consultation with
KROHNE service.

Zero point

000.0 ppm

Measurements beneath this adjustable value are displayed with this value.

Settings for auto-clean function

Designation

Settings

Description

Interval

000.0 h

Entry at which time intervals the auto-clean function is to be activated.
Recommendation: 000.5 h for compressed air, 002.0 h for water

Duration

000 sec

Rinse duration in seconds
Recommendation: 005 sec. for compressed air, 020 sec. for water

Hold time

000 sec

Waiting period after rinsing after which measurement can be engaged again. The
analogue outputs are frozen during the hold time. Recommendation: 030 sec.

Automatic

Yes
No

Yes

The auto-clean function is active

No

The auto-clean function is not active.