Settings for the display
Designation
Settings
Description
Units
lin
log
mV
pH
%
The selection possibilities depend on the respective parameter. The reference
magnitude between molecule or element can also be chosen during ISE measurement.
Modify
permitted
blocked
save data
load data
permitted
Entries and modifications are possible for this channel.
blocked
Entries and modifications are not possible
save data
Saves the current measuring channel settings.
load data
loads and uses the last saved channel settings.
System
Reset
Default
Test
Reset
The converter is restarted with the currently used parameters
Default
The converter is restarted with the factory settings. Existing customised settings are
lost.
Test
The converter switches to a test mode. Use this setting only in consultation with
KROHNE service.
Zero point
000.0 ppm
Measurements beneath this adjustable value are displayed with this value.
Settings for auto-clean function
Designation
Settings
Description
Interval
000.0 h
Entry at which time intervals the auto-clean function is to be activated.
Recommendation: 000.5 h for compressed air, 002.0 h for water
Duration
000 sec
Rinse duration in seconds
Recommendation: 005 sec. for compressed air, 020 sec. for water
Hold time
000 sec
Waiting period after rinsing after which measurement can be engaged again. The
analogue outputs are frozen during the hold time. Recommendation: 030 sec.
Automatic
Yes
No
Yes
The auto-clean function is active
No
The auto-clean function is not active.