Fixed allocations of analogue outputs
Designation
Settings
Description
Parameters
00:00:00:00
Allocation of 4 of the maximum 8 possible measuring channels (01 – 08) to the 4
analogue outputs. Attention, no channel may be indicated multiple times.
mA outputs
00:00:00:00
Free allocation of analogue outputs 01-04 to the measuring channel defined above
Baud rate RS232
02400
04800
09600
19200
38400
Factory setting 09600 baud.
Setting for data transmission between converter and PC. If possible, do not change the
baud rate on the converter.
If necessary, change the baud rate on your PC to the same baud rate as the converter.
Baud rate RS422
02400
04800
09600
19200
38400
Factory setting 19200 baud.
Setting for data transmission between converter and probe.
Do not change the baud rate, otherwise communication between converter and probe
may be disrupted.
In the case of doubt speak to the KROHNE service department beforehand.