KROHNE OPTISENS PAM 2080 EN User Manual

Page 50

background image

www.krohne.com

50

Compensation of constant disturbances (manual correction of cross-sensitivities )

Designation

Settings

Description

Correction value

+- 00.0 ppm

Here you can manually undertake compensations if disturbing ions disturb the
measurement by a constant amount.

Set a fixed correction value that is added to the measurement (+) or deducted from the
measurement (-).

After entering the correction value you must carry out an offset calibration using the
medium’s current measurement (measurement prior to entry of the correction value).

Calibrating the temperature sensor

Designation

Settings

Description

Temperature

025.0 °C

If the temperature sensor does not display the correct temperature, the temperature can
be calibrated here.
Implementation: enter the current temperature and then press Enter. Switch briefly into
display mode. Switch back to the menu and enter the same temperature once again.
Press Enter. The calibration temperature is saved.

Settings for calibration with standard addition (optional) only on Analyser DISCO-2

Designation

Settings

Description

Pump duration

00.1 min

Pump duration of a metering pump

Cond. volume

0.000 lt

Volume of conditioning medium which is to be set

Automatic

No
Yes

No
Manual activation of the standard addition

Yes
Automatic activation of the standard addition

Valve number

000

Sets which valve is to be controlled during standard addition

Back to measuring mode

Designation

Settings

Description

Menu

Parameters

Default

Parameters
Return to the measuring channel menu

Default
Restart with factory settings
The parameters used up to now are deleted.