beautypg.com

2 2-point-calibration – KROHNE OPTISENS PAM 2080 EN User Manual

Page 30

background image

www.krohne.com

30

6.2 2-Point-Calibration

2-point calibration takes place with two separate measuring solutions (standards) with known

concentrations.

pH- Redox electrodes:

To calibrate pH / Redox electrodes use ready-made standard buffer solutions that can also be procured via

the chemical / laboratory market.

Ion-sensitive probes (ISE):

To calibrate ion-sensitive probes do not use ready-made standard solutions from the laboratory market.

These are of no use for calibration, since as a synthetic measuring solution they never correspond to the

water matrix in which the online measurement is subsequently used. In practice, the resulting calibration

data are only utilisable to a limited extent and may sometimes lead to considerable deviations in

measurement.

Produce your own calibration standards

. This is not difficult, but the mistakes described above can be

avoided by doing so. Production of a customised standard solution is described below.

If possible, calibration solutions should be used that are within the prescribed measuring range for online

measuring. The use of concentrations at approx. 10-20% and approx. 80-90% of the expected measuring

range has proved effective (see example).

Example:

0 .. 50 mg/ NO3-N

measuring range set on PAM2080 (= mA output)

ca. 5 mg/l NO3-N

= 1st standard solution (10% measuring range)

ca. 45 mg/l NO3-N

= 2nd standard solution (90% measuring range)

Produce calibration standards with “your” specific water matrix.

For this you use the actual measuring water in which the online measurement is to function later.

This manual is related to the following products: