beautypg.com

KROHNE OPTISENS PAM 2080 EN User Manual

Page 33

background image

www.krohne.com

33

Now press the Cal button on the converter


Calibration is launched.

“Hold”, “Electrode active”, “Buffer”, “Run time” symbols displayed.

Calibration to the 1st standard lasts approx. 2 minutes. The symbols “Hold” and “Run time”
are displayed after that.

System waits for the 2nd standard.

Now place the immersion holder in the

2nd calibration standard.


First rinse the probe thoroughly again using clean water.