6.2.2 Carrying out 2-Point-Calibration

Select the measuring channel that you wish to calibrate.

Set the calibration type to 2-point calibration ("Calibration – Cal.method – 2-Point").

Now enter the determined concentrations (laboratory values) of the calibration solutions / pH buffer

for 1st buffer and 2nd buffer in the Cal menu ("Menu – Cal.menu – 2-Point-Calibration – 1

st

buffer /

2

nd

buffer").

Terminate the entry and return from the Cal menu via the parameter menu back to the measurement

display ("Menu – Cal.menu – Menu – Parameter")

Now place the immersion holder with the probe to be calibrated in the 1st calibration standard. First

clean the probe thoroughly using clean water.

The probe/immersion holder should be moved/swung in the calibration solution if possible.

Now press the Hold button on the converter.

Converter goes into system setting “Maintenance”.

Display symbol “Maintenance”.