Technical data, 7 permissible temperatures – KROHNE OPTITEMP TRA-TCA Plus EN User Manual
Page 72

7
TECHNICAL DATA
72
OPTITEMP TRA/TCA PLUS
www.krohne.com
01/2012 - 4000630301 - MA OPTITEMP TRA/TCA Plus R01
7.7 Permissible temperatures
7.7.1 Connection head
7.7.2 Unloaded thermowell and immersion tubes
The temperatures specified in the following tables are valid only for thermowells and immersion
tubes that do not project into flowing media or media under pressure. The maximum permissible
temperature is reduced under load.
Thermowells
Sheaths of measuring inserts made of mineral insulated cable
Temperature
transmitter
Connection head
Temperature range Maximum temperature
limiting component
Material
Display
without
Metal
no
-40...+100°C /
-40...+212°F
Gasket of connection head
cover and cable gland
without
Plastic
no
-40...+80°C /
-40...+176°F
Connection head material
with
Metal
no
-40...+85°C /
-40...+185°F
Temperature transmitter
with
Plastic
no
-40...+80°C /
-40...+176°F
Connection head material
with
Metal
Yes
-40...+70°C /
-40...+158°F
Display
DANGER!
If you are unsure as to whether the device can withstand the loads of the process when using an
insertion-type thermometer without thermowell, have a separate strength calculation done (e.g.
per DITTRICH or MURDOCK, ASME PTC 10.9-TW)! For further information contact the
manufacturer.
INFORMATION!
The manufacturer can provide an optional strength calculation for a fee.
Type of
thermowell
Material
Permissible temperature range
-T30
1.4571 / AISI 316 Ti
-200...+600°C / -328...+1112°F
1.4404 / AISI 316 L
-200...+600°C / -328...+1112°F
1.7335 / AISI F12
-200...+550°C / -328...+1022°F
1.0460 / C22.8
-200...+550°C / -328...+1022°F
All others
1.4571 / AISI 316 Ti
-200...+600°C / -328...+1112°F
1.4404 / AISI 316 L
-200...+600°C / -328...+1112°F
Type of sensor
Cable jacket material
Permissible temperature range
Pt100 RTD
1.4404 / AISI 316 L
-200...+600°C / -328...+1112°F
Thermocouple
Inconel
®
600
-40...+1100°C / -40...+1832°F
.backup.book Page 72 Tuesday, January 10, 2012 11:57 AM