Technical data – KROHNE OPTITEMP TRA-TCA Plus EN User Manual
Page 67

TECHNICAL DATA
7
67
OPTITEMP TRA/TCA PLUS
www.krohne.com
01/2012 - 4000630301 - MA OPTITEMP TRA/TCA Plus R01
7.6.2 Operating limits for one-piece, straight bar stock thermowells
TRA/TCA-TS36 and -TS37 with Ø 17 x 5 mm / 0.67 x 0.2"
TRA/TCA-TS52 with Ø 16 x 4.75 mm / 0.63 x 0.19"
TRA/TCA-TF55 with Ø 19 x 6.25 mm / 0.75 x 0.25"
Process connection
Material
Insertion length p
max
at +20°C / +400°C or
+68°F / +752°F, air
p
max
at +20°C / +400°C or
+68°F / +752°F, water /
vapour
[mm]
["]
[bar]
[psi]
[bar]
[psi]
Thread G½", G¾",
½
" NPT, ¾" NPT
1.4571 /
316 Ti,
1.4404 /
316 L
82
3.2
100 / 100
1450 / 1450
100 / 100
1450 / 1450
382
15.0
31.1 / 39.5
451 / 572
100 / 100
1450 / 1450
Process connection
Material
Insertion length p
max
at +20°C / +400°C or
+68°F / +752°F, air
p
max
at +20°C / +400°C or
+68°F / +752°F, water /
vapour
[mm]
["]
[bar]
[psi]
[bar]
[psi]
Thread G½", G¾",
½
" NPT, ¾" NPT
1.4571 /
316 Ti,
1.4404 /
316 L
100
3.9
100 / 100
1450 / 1450
100 / 100
1450 / 1450
300
11.8
44.8 / 56.7
649 / 822
100 / 100
1450 / 1450
Process connection
Material
Insertion length p
max
at +20°C / +400°C or
+68°F / +752°F, air
p
max
at +20°C / +400°C or
+68°F / +752°F, water /
vapour
[mm]
["]
[bar]
[psi]
[bar]
[psi]
Flanges ASME 1",
1½", 2" 150 lb
1.4404 /
316 L
100
3.9
15.9 / 6.5
231 / 94
15.9 / 6.5
231 / 94
300
11.8
15.9 / 6.5
231 / 94
15.9 / 6.5
231 / 94
Flanges ASME 1",
1½", 2" 300 lb
100
3.9
41.4 / 24.3
600 / 352
41.4 / 24.3
600 / 352
300
11.8
41.4 / 24.3
600 / 352
41.4 / 24.3
600 / 352
Flanges ASME 1",
1½", 2" 600 lb
100
3.9
82.7 / 48.6
1199 / 705
82.7 / 48.6
1199 / 705
300
11.8
62.0 / 48.6
899 / 705
82.7 / 48.6
1199 / 705
Flanges DN25, DN50
PN40
100
3.9
40.0 / 25.7
580 / 373
40.0 / 25.7
580 / 352
300
11.8
40.0 / 25.7
580 / 580
40.0 / 25.7
580 / 373
.backup.book Page 67 Tuesday, January 10, 2012 11:57 AM