Installation, 6 pre-installation requirements, 7 installation requirements sensor – KROHNE OPTISONIC 6300 Quickstart EN User Manual
Page 6
1
INSTALLATION
6
OPTISONIC 6300
www.krohne.com
07/2009 - 7.30959.24.00 - QS OPTISONIC 6300 R04 en
1.6 Pre-installation requirements
1.6.1 Environmental requirements
• Pollution degree 2
• Protection class I
• Humidity: 5...80 % RH
• Temperature: –40…+60°C / -40…+140°F operating and –50…+70°C / -58…+158°F storage
• Suitable for indoor and outdoor use and certified for operating up to an altitude of
2000 m / 6562 ft
• IP class 66/67
1.6.2 Installation requirements signal converter
• Allow 10…20 cm / 3.9…7.9" of space at the sides and rear of the signal converter to permit
free air circulation.
• Protect signal converter against direct solar radiation, install a sunshield if necessary.
• Signal converters installed in switchgear cabinets require adequate cooling, e.g. by fan or
heat exchanger.
• Do not expose the signal converter to intense vibration.
1.7 Installation requirements sensor
INFORMATION!
To assure a quick, safe and uncomplicated installation, we kindly request you to make provisions
as stated below.
CAUTION!
The device should be protected from corrosive chemicals or gases and dust / particles
accumulation.
INFORMATION!
To avoid measuring errors and malfunctioning of the flowmeter due to gas or air inclusions or an
empty pipe, please observe the following precautions.
CAUTION!
Since gas will collect at the highest point of a pipe, installation of the flowmeter at that location
should be avoided at all times. Also installation in a down going pipe should be avoided since a
completely filled pipe may not be guaranteed due to cascading effects. Additionally flow profile
distortion is possible.
CAUTION!
If you program the diameter, please note that you use the outer diameter of the pipe.
.book Page 6 Wednesday, August 19, 2009 8:37 AM