beautypg.com

Start-up – KROHNE OPTISONIC 6300 Quickstart EN User Manual

Page 40

background image

3

START-UP

40

OPTISONIC 6300

www.krohne.com

07/2009 - 7.30959.24.00 - QS OPTISONIC 6300 R04 en

• Mount the DOWN rail in such a way that the transducer is at the marked location.
• Grease all transducers, see "General mechanical installation".

Figure 3-9: Finding the location for the DOWN rail

1 Advised Distance as shown in menu X9.4
2 Measure the distance between the transducer and the end of the UP rail.
3 Determine and mark the location of the transducer of the DOWN rail: 3

3

3

3 = 1

1

1

1 - 2

2

2

2

INFORMATION!
It can be necessary to install the DOWN rail as shown below.

.book Page 40 Wednesday, August 19, 2009 8:37 AM