Electrical connections, 4 inputs and outputs, overview, 1 fixed, non-alterable input/output versions – KROHNE OPTISONIC 6300 Quickstart EN User Manual
Page 25

ELECTRICAL CONNECTIONS
2
25
OPTISONIC 6300
www.krohne.com
07/2009 - 7.30959.24.00 - QS OPTISONIC 6300 R04 en
2.4 Inputs and outputs, overview
2.4.1 Fixed, non-alterable input/output versions
This measuring transducer is available with various input/output combinations.
• The grey boxes in the tables denote unassigned or unused connection terminals.
• Connection terminal A+ is only operable in the basic input/output version.
CG-No. Connection terminals
A+
A
A-
B
B-
C
C-
D
D-
Basic input/output (I/O) standard
1 0 0
I
p
+ HART
®
passive
1
S
p
/ C
p
passive
2
S
p
passive
P
p
/ S
p
passive
2
I
a
+ HART
®
active
1
EEx-i inputs/outputs (I/Os) option
2 0 0
I
a
+ HART
®
active
P
N
/ S
N
NAMUR
2
3 0 0
I
p
+ HART
®
passive
P
N
/ S
N
NAMUR
2
2 1 0
I
a
active
P
N
/ S
N
NAMUR
C
p
passive
2
I
a
+ HART
®
active
P
N
/ S
N
NAMUR
2
3 1 0
I
a
active
P
N
/ S
N
NAMUR
C
p
passive
2
I
p
+ HART
®
passive
P
N
/ S
N
NAMUR
2
2 2 0
I
p
passive
P
N
/ S
N
NAMUR
C
p
passive
2
I
a
+ HART
®
active
P
N
/ S
N
NAMUR
2
3 2 0
I
p
passive
P
N
/ S
N
NAMUR
C
p
passive
2
I
p
+ HART
®
passive
P
N
/ S
N
NAMUR
2
1 function changed by reconnection
2 changeable