B.79 [SOURce:]VOLTage:PROTection:UNDer:STATE[:LEVel] COMMAND
VOLT:PROT:UND:STAT
Syntax:
Short Form: [SOUR:]VOLT:PROT:UND:STAT[:LEV]
Long Form: [SOURce:]VOLTage:PROTection:UNDer:STATe[:LEVel]
where
Description: Enables (1) or disables (0) the undervoltage protection level set using VOLT:PROT:UND. This
command saves having to change the protection level to some value outside of the operating range.
B.80 [SOURce:]VOLTage:PROTection:UNDer:STATE[:LEVel]? QUERY
VOLT:PROT:UND:STAT?
Syntax:
Short Form: [SOUR:]VOLT:PROT:UND:STAT[:LEV]?
Long Form: [SOURce:]VOLTage:PROTection:UNDer:STATe[:LEVel]?
Returns:
Description: Indicates whether the undervoltage protection level set using VOLT:PROT:UND is enabled (1)
or disabled (0).
B.81 [SOURce:]VOLTage:RANGe COMMAND
VOLT:RANG
Syntax:
Short Form: [SOURce:]VOLT:RANG
Where
Description: Controls the placement of the decimal point during remote interface readback of voltage, and
on the front panel VOLTS display when showing voltage. Display Ranges: 0 - Auto, 1 - 0.000, 2 -
00.00, 3 - 000.0, 4 - 0000.
B.82 [SOURce:]VOLTage:RANGe? QUERY
VOLT:RANG?
Syntax:
Short Form: [SOURce:]VOLT:RANG?
Long Form: [SOURce:]VOLTage:RANGe?
Returns
Description: Indicates the placement of the decimal point during remote interface readback of voltage, and
on the front panel VOLTS display when showing voltage. Display Ranges: 0 - Auto, 1 - 0.000, 2 -
00.00, 3 - 000.0, 4 - 0000.
B.83 STATus:CHANnel[:EVENt]? QUERY
STAT:CHAN?
Syntax:
Short Form: STAT:CHAN[:EVEN]?
Long Form: STATus:CHANnel[:EVENt]?
Returns: <16 bits> = the value of the Channel event register.
Description: Returns value of Channel event register. Each PFM within a chassis has three Channel registers
associated with it. Bits in the Event register are set whenever the associated condition occurs, and are
cleared only upon an explicit command, thus latching any conditions that occur. See Table B-2 for bit
assignments.
B.84 STATus:CHANnel:CONDition? QUERY
STAT:CHAN:COND?
Syntax:
Short Form: STAT:CHAN:COND?
Long Form: STATus:CHANnel:CONDition?
Returns: <16 bits> = the value of the Channel condition register.
Description: Returns the value of the Channel Condition Register. Each PFM within a chassis has three Chan-
nel registers associated with it. The Condition register reflects the instantaneous state of that PFM.
Bits in the Event register are set whenever the associated condition occurs, and are cleared only upon
an explicit command, thus latching any conditions that occur. See Table B-2 for bit assignments.
TABLE B-2. CHANNEL STATUS CONDITION,
ENABLE, AND EVENT REGISTER BITS
CONDITION
NU
PS
OV
LRV
UNR RRV
EPU
NU
OT
OP
NU OC VF
BIT
14-15
13
12
11
10
9
8
5 - 7
4
3
2
1
0
VALUE
16384-
32768
8192 4096 2048 1024
512
256
32-128
16
8
4
2
1
EPU EXTENDED PWR UNAVAILABLE
LRV LOCAL REVERSE VOLTAGE
NU -
NOT USED
OC
OVERCURRENT
OP
OVERPOWER
OT
OVERTEMPERATURE
OV
OVERVOLTAGE
PS
PROTECTION SHUTDOWN
RRV REMOTE REVERSE VOLTAGE
UNR UNREGULATED
VF
VOLTAGE FAULT