B.71 [SOURce:]VOLTage:MODE COMMAND
VOLT:MODE
Syntax:
Short Form: [SOUR:]VOLT:MODE
Long Form: [SOURce:]VOLTage:MODE
Description: Selects voltage mode.
B.72 [SOURce:]VOLTage:MODE? QUERY
VOLT:MODE?
Syntax:
Short Form: [SOUR:]VOLT:MODE?
Long Form: [SOURce:]VOLTage:MODE?
Returns:
Description: Returns the present mode: CURRent, POWer, VOLTage, RESistance, CONDuctance, OFF
B.73 [SOURce:]VOLTage:PROTection:OVEr COMMAND
VOLT:PROT:OVE
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVE
Long Form: [SOURce:]VOLTage:PROTection:OVEr
where
Description: Sets overvoltage protection limit. Overvoltage protection setpoint is stored in non-volatile memory;
value is restored upon power-up.
B.74 [SOURce:]VOLTage:PROTection:OVEr? QUERY
VOLT:PROT:OVE?
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVE?
Long Form: [SOURce:]VOLTage:PROTection:OVEr?
Returns
Description: Indicates overvoltage protection limit.
B.75 [SOURce:]VOLTage:PROTection:OVEr:STATe[:LEVel] COMMAND
VOLT:PROT:OVE:STAT
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVE:STAT[:LEV]
Long Form: [SOURce:]VOLTage:PROTection:OVEr:STATe[:LEVel]
where
Description: Enables (1) or disables (0) the overvoltage protection level set using VOLT:PROT:OVE. This
command saves having to change the protection level to some value outside of the operating range.
B.76 [SOURce:]VOLTage:PROTection:OVEr:STATe[:LEVel]? COMMAND
VOLT:PROT:OVE:STAT?
Syntax:
Short Form: [SOUR:]VOLT:PROT:OVE:STAT[:LEV]
Long Form: [SOURce:]VOLTage:PROTection:OVEr:STATe[:LEVel]
where
Description: Indicates whether the overvoltage protection level set using VOLT:PROT:OVE is enabled (1) or
disabled (0).
B.77 [SOURce:]VOLTage:PROTection:UNDer COMMAND
VOLT:PROT:UND
Syntax:
Short Form: [SOUR:]VOLT:PROT:UND
Long Form: [SOURce:]VOLTage:PROTection:UNDer
where
Description: Sets undervoltage protection limit. Undervoltage protection setpoint is stored in non-volatile mem-
ory; value is restored upon power-up.
B.78 [SOURce:]VOLTage:PROTection:UNDer? QUERY
VOLT:PROT:UND?
Syntax:
Short Form: [SOUR:]VOLT:PROT:UND?
Long Form: [SOURce:]VOLTage:PROTection:UNDer?
Returns:
Description: Indicates undervoltage protection limit.