B.39 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND
CURR
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]
Where
Description: Current Setting. Current setpoint is stored in non-volatile memory; value is restored upon power-up.
B.40 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? QUERY
CURR?
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]?
Returns
Description: Indicates Current Setpoint.
B.41 [SOURce:]CURRent:MODE COMMAND
CURR:MODE
Syntax:
Short Form: [SOUR:]CURR:MODE
Long Form: [SOURce:]CURRent:MODE
Description: Selects current mode.
B.42 [SOURce:]CURRent:MODE? QUERY
CURR:MODE?
Syntax:
Short Form: [SOUR:]CURR:MODE?
Long Form: [SOURce:]CURRent:MODE?
Returns:
Description: Returns the present mode: CURRent, POWer, VOLTage, RESistance, CONDuctance, OFF
B.43 [SOURce:]CURRent:PROTection[:LEVel] COMMAND
CURR:PROT
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV]
Long Form: [SOURce:]CURRent:PROTection[:LEVel]
where:
Description: Establishes current protection setpoint. Current protection setpoint is stored in non-volatile mem-
ory; value is restored upon power-up. This is a user-defined value which, if exceeded during opera-
tion, causes a FAULT condition, forcing the load to disengage.
B.44 [SOURce:]CURRent:PROTection[:LEVel]? QUERY
CURR:PROT?
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV]?
Long Form: [SOURce:]CURRent:PROTection[:LEVel]?
Returns:
Description: Indicates current protection setpoint.
B.45 [SOURce:]CURRent:PROTection:STATe COMMAND
CURR:PROT:STAT
Syntax:
Short Form: [SOUR:]CURR:PROT:STAT
Long Form: [SOURce:]CURRent:PROTection:STATe
where
Description: Enables (1) or disables (0) the current protection level set using CURR:PROT. This command
saves having to change the protection level to some value outside of the operating range.
B.46 [SOURce:]CURRent:PROTection:STATe? QUERY
CURR:PROT:STAT?
Syntax:
Short Form: [SOUR:]CURR:PROT:STAT?
Long Form: [SOURce:]CURRent:PROTection:STATe?
Returns:
Description: Indicates whether the current protection level set using CURR:PROT is enabled (1) or disabled
(0).