SERIES EL 063010

2-9

FIGURE 2-7. TYPICAL INTERCONNECTIONS

2.6.1

MAINS POWER

Mains power is connected at the rear panel of the Series EL Electronic Load (see Figure 2-3)
via the IEC 320-style recessed power inlet connector, which provides interface to the three-wire
safety line cord (supplied) via a polarized mating plug. (see PAR. 2.3 for mains power require-
ments).

It is recommended that the 1/4-inch bolt with wingnut and flat washers (supplied) at the top left
of the rear panel be connected to earth ground.

2.6.2

UNIT UNDER TEST (UUT) TO LOAD CONNECTION

WARNING!

Loose test connections between the UUT and the Series EL load can

lead to load instability, erratic operation and possible equipment

damage. In addition, loose connections or insufficiently sized

conductors can become very hot and cause burns.

2.6.2.1

POWER CONNECTIONS

Connections between the UUT and the Series EL Electronic Load are achieved via the (+) and
(–) LOAD terminals located on the rear panel.

The load is capable of drawing very high currents. Connections between the UUT and the load
should be of suitably large wire, sized to carry the maximum current. It is extremely important to
ensure all connections are tight and corrosion free. Regular inspection of these connections is
important. Inspection interval should be established based on frequency and type of use. Once
per week is recommended, however where cable movement or vibration is likely, more frequent
inspections are warranted.