B.23 [SOURce:]INPut:PROTection:SHORt[:STATe] COMMAND
INP:PROT:SHOR
Syntax:
Short Form: [SOUR:]INP:PROT:SHOR[:STAT]
Long Form: [SOURce:]INPut:PROTection:SHORt[:STATe]
Description: Applies maximum drive to FETs, effectively shorting the input.
B.24 [SOURce:]OUTPut[:STATe] COMMAND
OUTP
Syntax:
Short Form: [SOUR:]OUTP[:STAT]
where
Description: Used to engage or disengage the load from the UUT. When set to 1 the load engages, when set to
0 the load disengages.
B.25 [SOURce:]OUTPut[:STATe]? QUERY
OUTP?
Syntax:
Short Form: [SOUR:]OUTP[:STAT]?
Long Form: [SOURce:]OUTPut[:STATe]?
Returns
Description: Indicates whether load is engaged or disengaged from the UUT. When set to 1 the load is
engaged, when set to 0 the load is disengaged.
B.26 [SOURce:]OUTPut:PROTection:CLEar COMMAND
OUTP:PROT:CLE
Syntax:
Short Form: [SOUR:]OUTP:PROT:CLE
Long Form: [SOURce:]OUTPut:PROTection:CLEar
Description: Resets protection from overcurrent, overpower, etc.
B.27 [SOURce:]OUTPut:PROTection:SHORt[:STATe] COMMAND
OUTP:PROT:SHOR
Syntax:
Short Form: [SOUR:]INP:PROT:SHOR[:STAT]
Long Form: [SOURce:]INPut:PROTection:SHORt[:STATe]
Description: Applies maximum drive to FETs, effectively shorting the output.
B.28 [SOURce:]MEASure:ALL? QUERY
MEAS:ALL?
Syntax:
Short Form: [SOUR:]MEAS:ALL?
Long Form: [SOURce:]MEASure:ALL?
Returns: Measured values in the form 0.000 KW, 0.000 Amps, 0.000 Volts.
Description: Measures output power (Kilowatts), current (Amps), and voltage (Volts). Unit of measurement
included in the reply.
B.29 [SOURce:]MEASure:ALL2? QUERY
MEAS:ALL2?
Syntax:
Short Form: [SOUR:]MEAS:ALL2?
Long Form: [SOURce:]MEASure:ALL2?
Returns: Measured values in the form 0.000 0.000 0.000 (Kilowatts Amps Volts).
Description: Measures output power (Kilowatts), current (Amps), and voltage (Volts). Unit of measurement is omit-
ted from the reply.
B.30 [SOURce:]MEASure[:SCALar]:CURRent[:DC]? QUERY
MEAS:CURR?
Syntax:
Short Form: [SOUR:]MEAS[:SCAL]:CURR[:DC]?
Long Form: [SOURce:]MEASure[:SCALar]:CURRent[:DC]?
Returns: Measured values.
Description: Measures current. Measured returned as XXX Amps, where XXX is format determined by CUR-