B.40 [source:]conductance:range? query, B.43 [source:]current:mode command, B.44 [source:]current:mode? query – KEPCO EL Series Electronic Load Operator Manual P/N 243-1295 Firmware Version 3.12 (Unit Rev 2) User Manual
Page 67: B.45 [source:]current:protection[:level] command, B.46 [source:]current:protection[:level]? query, B.47 [source:]current:protection:state command, B.40, Source, B.41, B.42

SERIES EL 070110
B-7
B.40 [SOURce:]CONDuctance:RANGe? QUERY
COND:RANG?
Syntax:
Short Form: [SOUR:]COND:RANG?
Long Form: [SOURce:]CONDuctance:RANGe?
Returns
Description: Indicates the placement of the decimal point during remote interface readback of conductance,
and on the front panel KW display when showing conductance.
Display Ranges: 0 - Auto, 1 - 0.000, 2 - 00.00, 3 - 000.0, 4 - 0000.
B.41 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude] COMMAND
CURR
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]
Where
Description: Current Setting. Current setpoint is stored in non-volatile memory; value is restored upon power-up.
B.42 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]? QUERY
CURR?
Syntax:
Short Form: [SOUR:]CURR[:LEV][:IMM][:AMP]?
Long Form: [SOURce:]CURRent[:LEVel][:IMMediate][:AMPlitude]?
Returns
Description: Indicates Current Setpoint.
B.43 [SOURce:]CURRent:MODE COMMAND
CURR:MODE
Syntax:
Short Form: [SOUR:]CURR:MODE
Long Form: [SOURce:]CURRent:MODE
Description: Selects current mode.
B.44 [SOURce:]CURRent:MODE? QUERY
CURR:MODE?
Syntax:
Short Form: [SOUR:]CURR:MODE?
Long Form: [SOURce:]CURRent:MODE?
Returns:
Description: Returns the present mode: CURRent, POWer, VOLTage, RESistance, CONDuctance, OFF
B.45 [SOURce:]CURRent:PROTection[:LEVel] COMMAND
CURR:PROT
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV]
Long Form: [SOURce:]CURRent:PROTection[:LEVel]
where:
Description: Establishes current protection setpoint. Current protection setpoint is stored in non-volatile mem-
ory; value is restored upon power-up. This is a user-defined value which, if exceeded during opera-
tion, causes a FAULT condition, forcing the load to disengage.
B.46 [SOURce:]CURRent:PROTection[:LEVel]? QUERY
CURR:PROT?
Syntax:
Short Form: [SOUR:]CURR:PROT[:LEV]?
Long Form: [SOURce:]CURRent:PROTection[:LEVel]?
Returns:
Description: Indicates current protection setpoint.
B.47 [SOURce:]CURRent:PROTection:STATe COMMAND
CURR:PROT:STAT
Syntax:
Short Form: [SOUR:]CURR:PROT:STAT
Long Form: [SOURce:]CURRent:PROTection:STATe
where
Description: Enables (1) or disables (0) the current protection level set using CURR:PROT. This command
saves having to change the protection level to some value outside of the operating range.