B.15 DEBug:TREE:SEParator? QUERY

DEB:TREE:SEP?

Syntax:

Short Form: DEB:TREE:SEP?

Long Form: DEBug:TREE:SEParator?

Returns defines the column separator

Description: Returns the number corresponding to the selected column separator when producing a text

file of the command set using DEB:TREE:DISP command.

B.16 LOCK COMMAND

LOCK

Syntax:

Short Form: LOCK

Long Form: LOCK

where = OFF, EMERgency or LOCKed

Description: Controls functionality of front panel controls. When set to OFF, all front panel controls are fully

functional. When set to EMER, the front panel LOAD switch can only turn the load to the off state; all
other front panel controls except for the POWER switch are disabled. When set to LOCK, all front
panel controls except for the POWER switch are disabled.

B.17 LOCK? QUERY

LOCK?

Syntax:

Short Form: LOCK?

Long Form: LOCK?

Returns: = OFF, EMERgency or LOCKed

Description: Indicates state of front panel controls.

B.18 SENSe:CURRent:RANGe COMMAND

SENS:CURR:RANG

Syntax:

Short Form: SENS:CURR:RANG Long Form: SENSe:CURRent:RANGe
Where: = 0, 1, 2, 3 or 4

Description: Controls the placement of the decimal point during remote interface readback of current and

on the front panel AMPS display.
Display Ranges: 0 - Auto, 1 - 0.000, 2 - 00.00, 3 - 000.0, 4 - 0000.

B.19 SENSe:CURRent:RANGe? QUERY

SENS:CURR:RANG?

Syntax:

Short Form: SENS:CURR:RANG?

Long Form: SENSe:CURRent:RANGe?

Returns: = 0, 1, 2, 3 or 4

Description: Indicates the placement of the decimal point during remote interface readback of current and

on the front panel AMPS display.
Display Ranges: 0 - Auto, 1 - 0.000, 2 - 00.00, 3 - 000.0, 4 - 0000.

B.20 SENSe:VOLTage:RANGe COMMAND

SENS:VOLT:RANG

Syntax:

Short Form: SENS:VOLT:RANG

Long Form: SENSe:VOLTage:RANGe

Where: = 0, 1, 2, 3 or 4

Description: Controls the placement of the decimal point during remote interface readback of voltage, and

on the front panel VOLTS display.
Display Ranges: 0 - Auto, 1 - 0.000, 2 - 00.00, 3 - 000.0, 4 - 0000.

B.21 SENSe:VOLTage:RANGe? QUERY

SENS:VOLT:RANG?

Syntax:

Short Form: SENS:VOLT:RANG? Long

Form:

SENSe:VOLTage:RANGe?

Returns: = 0, 1, 2, 3 or 4

Description: Indicates the placement of the decimal point during remote interface readback of voltage, and

on the front panel VOLTS display.
Display Ranges: 0 - Auto, 1 - 0.000, 2 - 00.00, 3 - 000.0, 4 - 0000.

B.22 [SOURce:]INPut[:STATe] COMMAND

INP

Syntax:

Short Form: [SOUR:]INP[:STAT]

Long Form: [SOURce:]INPut[:STATe]

where = 0 or OFF (load disengaged) or 1 or ON (load engaged)

Description: Used to engage or disengage the load from the UUT. When set to 1 the load engages at the pro-

grammed mode and corresponding setpoint value; when set to 0 the load disengages.