29 device status, Device status – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual
Page 74
8.29 Device Status
The device status register holds general status information. Each set bit represents a status
value. There are a limited number of standardized predefined status values that all devices will
support. These predefined status values are contained in the lower register. The upper register
is reserved for device specific status values.
Bit
Category
Description
0
Alarm
Sensor high alarm
1
Warning
Sensor high warning
2
Warning
Sensor low warning
3
Alarm
Sensor low alarm
4
Warning
Sensor calibration warning
5
Alarm
Sensor malfunction
6-7
N/A
Reserved
8
Status
Power management disabled
9
Status
Device off line
10
Alarm
Device hardware reset occurred
11
Alarm
Device malfunction
12
Status
No external power.
13
Warning
Low battery – battery capacity < 5%
14
Warning
Low memory – data log memory capacity < 5%
15
N/A
Reserved
16-31
N/A
Available for device-specific status
• Bits 0-7 of the device status register are reserved for sensor status. These bits are
the logical OR of bits 0-7 of the sensor status register in each sensor connection.
• Bits 8-15 of the device status register are reserved for common device status. Any bit
in this range that is not applicable to a device will be set to zero.
• Bits 16-31 of the device status register are available for device-specific status. Any
bit in this range that is not utilized by a device will be set to zero.
74