8.19 Logged Record Registers
Register
XXXX
Size
(registers
Mode
(R/W)
Data
Type
Description
9600-601
2
R1
ulong Number of logged records
9602-603
2
R1/W3
ulong Requested log record number
9604-606
3
R1
time Time stamp for record
9607-608
2
R1
float Parameter 1 measured value
9609
1
R1
ushort Parameter 1 data quality ID
96010-696
29 x 3
R1
---
Parameters 2 – 30
(30 is max allowed #log parameters)
8.20 Register Map Template ID
This is a number that represents the version of the reserved register map supported by the
device. This provides for the modification of the reserved register map at a future date and
detection of this difference by the Master. This register is commonly referred to as the System
Interface Specification Level or (SIS Level).