5. Modbus Registers—Conductivity Sensor

5.1 Header Registers—Conductivity

The sensor map data register offset points to the first register in the sensor data header block.
The current value for this sensor data register map version is 1.

Register

Size

(registers)

Mode & Access Level

(R/W)

Data

Type

Description

0501

1

R1

ushort

Sensor ID = 35

0502

2

R1

ulong

Sensor serial number

(same as probe)

0504

1

R1

16 bits

Sensor status

0505

3

R1/W4

time

Last factory calibration

0508

3

R1/W4

time

Next factory calibration

(0 = none required)

0511

3

R1

time

Last user calibration

0514

3

R1/W2

time

Next user calibration

(0 = none required)

0517

1

R1

ushort

Warm-up time in

milliseconds = 1200

0518

1

R1

ushort

Fast sample rate =

1200 milliseconds

0519

1

R1

ushort

Number of sensor

parameters (N) = 7

0520

1

R1/W3

ushort

Alarm/warning parameter

number (1 – N, default = 1)

0521

1

R1/W3

16 bits

Alarm and warning enable

bits (default = 0)

Bit 0 = High alarm enabled

Bit 1 = High warning enabled

Bit 2 = Low warning enabled

Bit 3 = Low alarm enabled

Bit 4 = Sensor calibration warning

0522

2

R1/W3

float

High alarm set value

(default = 0.0)

36