16 communication control registers, 17 sensor connection registers, Communication control registers – In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual
Page 70: Sensor connection registers

8.16 Communication Control Registers
8.17 Sensor Connection Registers
Register
XXXX
Size
(registers
Mode
(R/W)
Data
Type
Description
9300
1
R1
ushort Max Sensor Connections (1-32)
9301-302
2
R1
32 bits Sensor Connection Status
Sensor Map Registers
RDO Sensor
9303
1
R1
ushort Connection 1 sensor id
9304
1
R1
16 bits Connection 1 sensor status
9305
1
R1/W2
ushort Connection 1 sensor command
9306
1
R1
ushort Connection 1 sensor data register map version
9307
1
R1
ushort Connection 1 sensor data register offset
Conductivity Sensor
9308
1
R1
ushort Connection 2 sensor id
9309
1
R1
ushort Connection 2 sensor status
9310
1
R1
ushort Connection 2 sensor command
Register
XXXX
Size
(registers
Mode
(R/W)
Data
Type
Description
9200
1
R/W
ushort
Device Address (1-247, default = 1)
9201
1
R/W
ushort
Serial Communication Configuration
9202
1
R/W
ushort
EOM timeout (1000-15000 ms, default = 1000)
9203
1
R/W
ushort
EOS timeout (5000-60000 ms, default = 5000)
9204
1
R
ushort
Max allowed baud rate id (0-7)
9205
1
R
ushort
Max Message/Response size (bytes)
9206-207
2
R/W
ulong
Good message counter
9208
1
R/W
ushort
Bad message counter
9209
1
R/W
ushort
Exception response counter
70