In-Situ Aqua TROLL 400 Modbus and SDI-12 Reference Guide User Manual
Page 16

Name
Command
Response
Start Concurrent
Start Concurrent CRC
Send Data
Additional Data
aC!
aCC!
aD0!
aD1! … aD9!
a004nn
nn parameters will be available for reading by the Send
Data command within 4 seconds. No service request
will be sent when the parameters are ready. The
number of parameters returned is determined by the
SDI-12 configuration file in the same manner as a Start
Measurement command.
a
a
The number and type of parameters returned is
determined by the SDI-12 configuration file in the same
manner as a Start Measurement command.
a
a
The number and type of parameters returned is
determined by the SDI-12 configuration file in the same
manner as a Start Measurement command.
Additional Concurrent
Additional with CRC
Send Data
Additional Data
aC1! … aC9!
aCC1! … aCC9!
aD0!
aD1! … aD9!
a00000
No additional concurrent measurements are started by
the device.
a
a
No values are returned after an additional Concurrent
Measurement command.
Continuous Measurement
Continuous with CRC
aR0! … aR9!
aRC0! … aRC9!
a
No values shall be returned after a Continuous
Measurement command.
16