beautypg.com

Precautions before operation – GW Instek GOS-6200 User Manual

Page 7

background image

GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 8 ⎯

3.PRECAUTIONS BEFORE OPERATION

3-1.Unpacking the Oscilloscope

The product has been fully inspected and tested before shipping from the

factory. Upon receiving the instrument, please unpack and inspect it to

check if there is any damages caused during transportation. If any sign of

damage is found, notify the bearer and/or the dealer immediately.

3-2.Checking the Line Voltage

The oscilloscope can be applied any kind of line voltage shown in the

table below. Before connecting the power plug to an AC line outlet, make

sure the voltage selector of the rear panel is set to the correct position

corresponding to the line voltage. It might be damaged the instrument if

connected to the wrong AC line voltage.

WARNING. To avoid electrical shock the power cord

protective grounding conductor must be connected to ground.

When line voltages are changed, replace the required fuses shown as below:

Line voltage

Range

Fuse

Line voltage

Range

Fuse

100V
120V

90-110V

108-132V

T 1A250V

230V

207-250V

T 0.4A250V

WARNING. To avoid personal injury, disconnect the power

cord before removing the fuse holder.

GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 9 ⎯

3-3.Environment

The normal ambient temperature range of this instrument is from 0

° to

40

°C (32° to 104°F). To operate the instrument over this specific

temperature range may cause damage to the circuits.

Do not use the instrument in a place where strong magnetic or electric

field exists as it may disturb the measurement.

3-4.Equipment Installation, and Operation

Ensure there is proper ventilation for the vents in the oscilloscope case. If

the equipment is used not according to the specification, the protection

provided by the equipment may be impaired.

3-5.CRT Intensity

To prevent permanent damage to the CRT phosphor, do not make the

CRT trace brighten excessively or leave the spot stay for an unreasonably

long time.

3-6.Withstanding Voltages of Input Terminals

The withstanding voltages of the instrument input terminals and probe

Input terminals are shown in the following table. Do not apply voltages

higher than these limits.

Input terminal

Maximum input voltage

CH1, CH2, inputs

400V (DC + AC peak)

EXT TRIG input

400V (DC + AC peak)

Probe inputs

600V (DC + AC peak)

Z AXIS input

30V (DC + AC peak)

CAUTION. To avoid damaging the instrument, do not apply input

voltages of the frequency over 1 kHz to the instrument.