GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 16 ⎯

(13)CH1 VOLTS/DIV.

(14)CH2 VOLTS/DIV– Control knob for channel 1/channel 2 has double

function.

Turning the knob clockwise to increase the sensitivity in 1-2-5 sequence

and turning it in the opposite direction (CCW) to decrease. The

available range is from 2mV/div up to 5V/div. The knob is

automatically switched inactive if the related channel is switched off.

The deflection coefficients and additional information regarding the

active channels are displayed in the readout.

ie. “CH1=deflection coefficient, input coupling”. The “=” symbolizes

calibrated measuring conditions and is replaced by the “>” symbol in

uncalibrated conditions.

(15)CH1-VAR.

(16)CH2-VAR — Pushbutton with double function.

CH1/CH2

Pressing briefly the CH1(CH2) button to set the cannel 1 (channel 2) of

the instrument on, the deflection coefficient will be displayed in the

readout indicating the current conditions (“CH1…”/ “CH2…”).

VAR

Pressing and holding the pushbutton to select the VOLTS/DIV function

between attenuator and vernier (variable). The current setting is

displayed by the “>” symbol in the readout.

After switching on the VAR, turn the VOLTS/DIV control knob

counterclockwise to reduce the signal height, and the deflection

coefficient becomes uncalibrated.

GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 17 ⎯

(17)CH1 AC/DC.

(18)CH2 AC/DC

Pressing the pushbutton briefly to switch over from AC (~ symbol) to

DC (= symbol) input coupling. The setting is displayed in the readout

with the deflection coefficient.

(19)CH1 GND– P×10

(20)CH2 GND – P×10 –Pushbutton has two functions.

GND

Each time when the pushbutton is pressed briefly, the input of the

vertical amplifier is grounded. It is displayed in the readout as an earth

(ground) symbol “ ”.

P×10

Pressing and holding the pushbutton to select the indicated deflection

coefficient of the channel displayed in the readout between 1:1 and 10:1.

The probe factor of 10:1 is displayed in the readout with the probe

symbol in front of channel indication (e.g. “P10”, CH1) When proceed

cursor voltage measurement, the probe factor will be automatically

included. The symbol must not be activated unless a 10:1 attenuator

probes are used.