GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 54 ⎯

1. Set the VOLTS/DIV and VAR controls to provide an exact five-

division vertical display.

2. Use the vertical POSITION control to control the negative amplitude

of the signal on the 0% reference line and the positive amplitude on

the 100% reference line.

3. Increase the TIME/DIV setting to stretch out the rising edge of the

waveform as much as possible to improve the cursor placement

accuracy.

4. Use the C1-POSITION control to align the cursor 1 to the rising

edge at the point where it crosses the 10% reference graticule line.

Then use the C2-POISITION control to align the cursor 2 to the

point where the rising edge crosses the 90% graticule line and read

the rise time displayed in the CRT readout.

(e).Typical 1/△T cursor function for frequency

measurement.

When the two cursors are superimposed at

two edge points of the one period waveform

by the C1-POSITION and C2-POSITION

controls, the measurement value is displayed

in frequency units on the upper side of the

screen.

(f).Typical △ T%(Time difference percentage)

cursor function for duty-cycle measurement of

square waveform.

A time difference percentage measurement is

done by first setting a reference for the full

cycle of waveform period: 5div=100%.

GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 55 ⎯

(g).Typical △ Θ cursor function for phase

measurement.

A phase measurement is done by first setting

a reference for the full 360

o

waveform

period:

5div=360

o

.

NOTE. When the VOLTS/DIV or the TIME/DIV controls are in

uncalibrated setting, the V and T measurement values will be

displayed with divisions.

When the vertical mode is set to the ADD mode, and the CH1 and CH2

VOLTS/DIV controls are set to different scales, the V measurement

values will be displayed with divisions.