GW Instek GOS-6200 User Manual
Page 30
GOS-6200 OSCILLOSCOPE
USER MANUAL
⎯ 54 ⎯
1. Set the VOLTS/DIV and VAR controls to provide an exact five-
division vertical display.
2. Use the vertical POSITION control to control the negative amplitude
of the signal on the 0% reference line and the positive amplitude on
the 100% reference line.
3. Increase the TIME/DIV setting to stretch out the rising edge of the
waveform as much as possible to improve the cursor placement
accuracy.
4. Use the C1-POSITION control to align the cursor 1 to the rising
edge at the point where it crosses the 10% reference graticule line.
Then use the C2-POISITION control to align the cursor 2 to the
point where the rising edge crosses the 90% graticule line and read
the rise time displayed in the CRT readout.
(e).Typical 1/△T cursor function for frequency
measurement.
When the two cursors are superimposed at
two edge points of the one period waveform
by the C1-POSITION and C2-POSITION
controls, the measurement value is displayed
in frequency units on the upper side of the
screen.
(f).Typical △ T%(Time difference percentage)
cursor function for duty-cycle measurement of
square waveform.
A time difference percentage measurement is
done by first setting a reference for the full
cycle of waveform period: 5div=100%.
GOS-6200 OSCILLOSCOPE
USER MANUAL
⎯ 55 ⎯
(g).Typical △ Θ cursor function for phase
measurement.
A phase measurement is done by first setting
a reference for the full 360
o
waveform
period:
5div=360
o
.
NOTE. When the VOLTS/DIV or the TIME/DIV controls are in
uncalibrated setting, the △V and △T measurement values will be
displayed with divisions.
When the vertical mode is set to the ADD mode, and the CH1 and CH2
VOLTS/DIV controls are set to different scales, the △V measurement
values will be displayed with divisions.