GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 20 ⎯

window segment decreases when the DELAY time coefficient is set to

a lower value (higher time deflection speed).

For better reading, the vertical position of the DELAY time base trace

position can be shifted (please note TRACE SEP (11)).

DELAY

In the DELAY time base mode, the display of the MAIN traces, the

window sector and the MAIN time coefficient will disappear from the

readout. As the trace separation is no longer required under the

circumstances, the function would be switched off too. Consequently,

only the DELAY time coefficient is displayed by the readout.

(23)H POSITION

The control knob enables a horizontal position shift of the signals. In

combination with ×10 MAG the function makes it possible to shift any

part of the signal on the screen.

(24)×10 MAG—SETUPS LOCK— Pushbutton has double function and

associated MAG LED.

Each time when this pushbutton is pressed, the MAG LED located

above will be switch on or off. If the MAG LED is lighted, the signal

display in all time base modes will be expanded 10 folds and

consequently only a tenth part of the signal curve is visible. The

interesting part of the signal can be made visible with the aid of the H.

POSITION control.

SETUPS LOCK

Pressing and holding the pushbutton, then switch the panel setups lock

function on or off. To avoid unintentional touch of the setting, the

feature is extremely useful for long term and repetitive measurements

that need to be performed under the same test condition of the

oscilloscope setting.

GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 21 ⎯

(25)X-Y-VAR – Pushbutton with double function.

X-Y

Set to X-Y mode to select three functions sequentially by pressing the

button briefly.

VAR

Pressing and holding the pushbutton to select the TIME/DIV (21)

control knob function between time base switch and vernier (variable).

The variable function is activated in the MAIN time base only.

After switching on the VAR, the time deflection coefficient is still

calibrated until further adjustments are made. Turn the TIME/DIV (21)

control knob counter clockwise to increase the time deflection

coefficient (reduce the deflection speed) and the deflection coefficient

becomes uncalibrated. Instead of “A=10μs”, the readout then displays

“A>10μs” indicating the uncalibrated condition. This setting is stored

if the instrument is switched to ALT or DELAY time base mode.

Switch off the VAR by pressing and holding the pushbutton of time

base mode again, then set the time deflection coefficient back into the

calibrated condition.