GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 42 ⎯

4. Set both CH1 and CH2 COUPLING to GND.

5. Use the CH1 and CH2 POSITION controls to align both traces on the

center graticule.

6. Open the CH2 INV by pressing and holding the pushbutton.

7. Set to the ADD mode by pressing the ALT/CHOP/ADD pushbutton

briefly.

8. Set both CH1 and CH2 COUPLING to DC.

9. The figure 5-4 below shows a satisfactory display. The display will

show a flat trace located on the center gracticule that confirms the

channel balance and ADD offset function.

Figure 5-4 Display 2

10. Set to the CHOP mode by pressing the ALT/CHOP/ADD pushbutton

briefly.

11. Turn off the CH2 INV by pressing and holding the pushbutton.

GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 43 ⎯

5-5.Basic Operation

Displaying CH1 or CH2

To display the signal from a signal channel, pressing briefly the CH1 or

CH2 pushbutton to set the oscilloscope to channel 1 or channel 2.

Displaying CH1 and CH2

To display both signals at the same time, proceed the following steps:

1.Set the CH1 and CH2 on. The figure 5-5 below shows two synchronous

waveforms in the both modes.

2.Adjust the CH1 or CH2 POSITION control to position the two

waveforms.

3.Set the ALT/CHOP/ADD button to CHOP mode if the waveforms are

flickering.

Figure 5-5 Both typical waveforms