Horizontal controls:

The horizontal controls select the time base operation mode and adjust the

horizontal scale, position and magnification of the signal.

(21)TIME/DIV– Control knob with double function.

Turning the knob clockwise to reduce the deflection coefficient in a

1-2-5 sequence and turning it in the opposite direction (CCW) to

increase. The time coefficient(s) will be displayed in the readout.

In MAIN time base (MTB) mode, time deflection coefficients between

0.5s/div and 20ns/div can be chosen in 1-2-5 sequence, if the ×10

MAG function is not activated.

During alternate (ALT) and DELAY time base (DTB) operation, the

control knob changes the DELAY time base setting in 1-2-5 sequence.

The available deflection coefficient range is from 50ms/div up to

20ns/div (without ×10 MAG), but the availability depends on the

MAIN time base setting. The internal control of the oscilloscope

prevents the DELAY time deflection coefficient from becoming higher

than the MAIN deflection coefficient, as such an operation condition

would make no sense.

GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 19 ⎯

(22)MAIN/ALT/DELAY— Pushbutton for time base mode selection.

The instrument contains two-time base designated MAIN and DELAY.

With the aid of the DELAY time base, signal parts displayed by the

MAIN time base can be expanded in X-direction. The expansion ratio

depends on the time deflection coefficient ratio of both time bases (ie.

“MTB=0.1ms”, “DTB=1μs”=100:1). With higher expansion ratio the

DELAY time base trace intensity reduces. Each time when press the

pushbutton briefly, the time base mode changes in the sequence of

MAIN-ALT-DELAY-MAIN. The actual setting is displayed in the

readout.

MAIN

The TIME/DIV control knob is operated only under the MAIN time

base mode. The readout then displays the main time coefficient alone.

The time base setting for this condition will be stored if the time base

mode is changed.

ALT

If the alternate time base mode is selected, the TIME/DIV knob only

controls the DELAY time base switch. The alternate time base mode is

a sub-function of the DELAY time base mode and both time base

traces can be displayed simultaneously. Consequently the readout can

display both time deflection coefficient. A window sector which

indicates part of signal is also visible on the MAIN trace and is

displayed by the DELAY time base.

The window segment can be shifted horizontally by the DELAY TIME

control continuously. The difference between the beginning of both the

MAIN time base trace and the window sector shows the delay time.

The information is also displayed in the readout with an approximate

value (e.g. “DLY=0.125ms”) related to the calibrated MAIN time

coefficient (uncalibrated i.e. “DLY>0.125ms”). The width of the