GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 40 ⎯

7. Observe the displayed waveform and compare them with the waveforms

shown in figure 5-2. If either probe needs to be adjusted, proceed the

step 8. If either probe does not need to be adjusted, proceed the

“Function Check”.

Figure 5-2 Typical Compensation Waveform

8.Adjust the probe by using a small insulated screwdriver. Slowly rotate

the adjustment control until the probe is properly compensated.

GOS-6200 OSCILLOSCOPE

USER MANUAL

⎯ 41 ⎯

5-4.Function Check

When you start to check the operation of your oscilloscope, proceed the

following instruction:

1. Install the ×10 probes onto CH1 and CH2 inputs.

2. Connect the probe tips to the CAL test point of the oscilloscope.

3. Set the oscilloscope controls to display both channels:

VERTICAL:

VOLTS/DIV

1V

COUPLING

DC

ALT/CHOP/ADD

CHOP

HORIZONTAL:

MODE

MTB

TIME/DIV

0.5ms

TRIGGER:

MODE

ATO

SOURCE

CH1 or CH2

COUPLING

AC

SLOPE

The figure 5-3 below illustrates a satisfactory display. The waveform

should be approximately 2Vp-p at a frequency of 1kHz that confirms

the vertical and horizontal deflection function of the oscilloscope.

Figure 5-3 Display 1