GSP-930 Quick Start Guide

54

Tracking Generator (Optional)

Frequency Range

100 kHz to 3 GHz

Output Power

-50 dBm to 0 dBm in 0.5 dB steps

Absolute Accuracy ± 0.5 dB

@160 MHz, -10 dBm,
Source attenuation 10 dB,
20 to 30°C

Output Flatness

Referenced to 160 MHz, -10 dBm

100 kHz to 2 GHz

± 1.5 dB

2 GHz to 3 GHz

± 2 dB

Output Level
Switching
Uncertainty

± 0.8 dB

Referenced to -10 dBm

Harmonics

< -30 dBc

Typical, output level = -10
dBm

Reverse Power

+30 dBm max.

Connector type

N-type female

Impedance

50 ohm

Nominal

Output VSWR

< 1.6:1

300 kHz to 3 GHz, source
attenuation ≥ 12 dB

USB Power Sensor (Optional)

Type

Average power sensor Model: PWS-06

Interface to Meter USB cable to GSP930 Front-Panel USB Host

Connector Type

N-type male, 50 ohm nominal

Input VSWR

1.1: 1

Typical

1.3: 1

Max

Input Frequency

1 to 6200 MHz

Sensing Level

-32 to +20 dBm

Max. Input Damage
Power

≥ 27 dBm