beautypg.com

GW Instek GSP-930 Quick start guide User Manual

Page 51

background image

APPENDIX

49

1 dB Gain Compression

Total Power at 1st
Mixer

> 0 dBm

Typical

;Fc ≥ 50 MHz;

preamp. off

Total Power at the
Preamp

> -22 dBm

Typical

;Fc ≥ 50 MHz;

preamp. on

mixer power level
(dBm)= input power
(dBm)-attenuation
(dB)

Displayed Average Noise Level (DANL)

Preamp off

0 dB attenuation; RBW 10 Hz; VBW 10 Hz; span
500 Hz; reference level = -60dBm; trace average
≥ 40

9 kHz to 100 kHz < -93 dBm,

Nominal

100 kHz to 1
MHz

< -90 dBm - 3 x (f/100
kHz) dB

1 MHz to 10 MHz < -122 dBm

10 MHz to 3 GHz < -122 dBm

Preamp on

0 dB attenuation; RBW 10 Hz; VBW 10Hz; span
500 Hz; reference level = -60dBm; trace average
≥ 40

100 kHz to 1
MHz

< -108 dBm - 3 x (f/100
kHz) dB

Nominal

1 MHz to 10 MHz < -142 dBm

10 MHz to 3 GHz < -145 dBm + 3 x (f/1

GHz) dB

Level Display Range

Scales

Log, Linear

Units

dBm, dBmV, dBuV, V, W

Marker Level
Readout

0.01 dB

Log scale

0.01 % of reference level Linear scale

Level Display
Modes

Trace, Topographic,
Spectrogram

Single / split
Windows

Number of Traces 4

Detector

Positive-peak, negative-
peak, sample, normal,
RMS(not Video)

Can be setup for each
trace separately

Trace Functions

Clear & Write, Max/Min
Hold, View, Blank, Average