AM8060 Precision Thermometer

User’s Guide

8/43

probe/probes and temperature difference. This information is

shown in 4 lines, as seen below:

Line 1 is the instrument interface status, including USB flash

drive, wireless communication status and USB to computer

interface status. If the “Udr”is shown in Line 1, a USB flash

drive is connected to the USB host interface; if “ZB-xx (where

“xx” are numbers) displays on Line 1 as shown above, it means

the Zigbee wireless data capturing is working; if the display

shows“USB”on Line 1, then AM8060 is connected via its

USB client interface.

Line 2, which is indicated with ‘t1’, shows the measured

temperature of Probe 1 (S1). The“SPRT” at the end of the

Line 2 tells the type of probe of Probe 1 (S1). The AM8060

supports two types of probes. “SPRT” indicates a 25Ω or 100 Ω

standard platinum resistance thermometers. “IPRT” indicates a

100Ω industrial platinum resistance thermometer. The “C”

shown at the end of Line 2 indicates that the temperature is in

Celsius.

Line 3, which is indicated with ‘t2’, shows the measured

temperature of Probe 2 (S2).